What information needs to be provided to the Competent Authorities prior to initiation of a clinical study with a product without a CE-Mark or outside of the labelling of a CE-Marked product? What basic information about the medical device needs to be maintained on-file for inspection by the Competent Authorities?

 

Council Directive 90/385/EEC

(5 September 2007)

 

ANNEX 6

STATEMENT CONCERNING DEVICES INTENDED FOR SPECIAL PURPOSES

 

1.     The manufacturer or his authorized representative established within the Community shall draw up for custom-made devices or for devices intended for clinical investigations the statement comprising the elements stipulated in section 2.

 

2.         The statement shall comprise the following information:

 

2.1.      For custom-made devices: -/-

  

2.2. For devices intended for clinical investigations covered in Annex 7:

 

  • data allowing the devices in question to be identified,
  • the clinical investigation plan,
  • the investigator's brochure,
  • the confirmation of insurance of subjects,
  • the documents used to obtain informed consent,
  • a statement indicating whether or not the device incorporates, as an integral part, a substance or human blood derivative referred to in Section 10 of Annex 1,
  • the opinion of the ethics committee concerned and details of the aspects covered by its opinion,
  • the name of the duly qualified medical practitioner or other authorised person and of the institution responsible for the investigations,
  • the place, date of commencement and duration scheduled for the investigations,
  • a statement affirming that the device in question complies with the essential requirements apart from the aspects constituting the object of the investigations and that, with regard to these aspects, every precaution has been taken to protect the health and safety of the patient.

 

3.         The manufacturer shall undertake to keep available for the competent national authorities:

 

3.1.      For custom-made devices, -/-

 

3.2.      For devices intended for clinical investigations, the documentation shall also contain:

 

  • a general description of the product and its intended use,
  • design drawings, manufacturing methods, in particular as regards sterilization, and diagrams of parts, sub-assemblies, circuits, etc.,
  • the descriptions and explanations necessary for the understanding of the said drawings and diagrams and of the operation of the product,
  • the results of the risk analysis and a list of the standards laid down in Article 5, applied in full or in part, and a description of the solutions adopted to satisfy the essential requirements of the Directive where the standards in Article 5 have not been applied,
  • if the device incorporates, as an integral part, a substance or human blood derivative referred to in Section 10 of Annex 1, the data on the tests conducted in this connection which are required to assess the safety, quality and usefulness of that substance, or human blood derivative, taking account of the intended purpose of the device,
  • the results of the design calculations, checks and technical tests carried out, etc.

 

The manufacturer shall take all necessary measures to see that the manufacturing process ensures that the products manufactured conform to the documentation referred to in 3.1 and in the first paragraph of this section.

The manufacturer may authorize the evaluation, by audit where necessary, of the effectiveness of these measures.

 

4. The information included in the declarations covered by this Annex shall be kept for a period of at least 15 years from the date of manufacture of the last product.

 

5.         For custom-made devices, -/-

  ISO 14155:2011 Weblink  
 

Clinical investigation of medical devices for human subjects - Good clinical practice
Document available as of: 2011-01-21

 

ISO 14155:2011 addresses good clinical practice for the design, conduct, recording and reporting of clinical investigations carried out in human subjects to assess the safety or performance of medical devices for regulatory purposes.

 

The principles set forth in ISO 14155:2011 also apply to all other clinical investigations and should be followed as far as possible, depending on the nature of the clinical investigation and the requirements of national regulations.

 

ISO 14155:2011 specifies general requirements intended to protect the rights, safety and well-being of human subjects, ensure the scientific conduct of the clinical investigation and the credibility of the results, define the responsibilities of the sponsor and principal investigator, and assist sponsors, investigators, ethics committees, regulatory authorities and other bodies involved in the conformity assessment of medical devices.

 